Hydac HLB 1400 Instrukcja Użytkownika Strona 8

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 18
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 7
Instruction Manual HYDACLab
03.03.2008
8
8. Technical Data
Operating Conditions
Relative humidity 0 .. 100 % of saturated concentration
Temperature –25°C .. + 100°C
Viscosity 1 cSt .. 1,000 cSt
Dielectric constant (ε
R
) 1.5 .. 10
Operating pressure < 50 bar
Pressure resistance 600 bar
Fluid flow velocity < 5 m/s
Output data – Humidity measurement
Output signal 4 mA to 20 mA (corresponds to 0 .. 100 %)
Calibration accuracy
± 2 % of FS max.
Accuracy for measurements in media
± 3 % of FS (typical) *
Output data – Temperature measurement
Output signal 4 .. 20 mA ( corresponds to – 25 .. + 100°C)
Accuracy
± 3 % of FS max.
Output data – Relative change in viscosity and dielectric constant (ε
R
)
Output signal 12 mA ± 8 mA (corresponds to ± 30 % of
IV)
Accuracy See below **
Switching Output
Signal 1 (NC) PNP-transistor switching output, 0.5 A max.
current draw; switching level: (U
B
– 4 V)
Preset warning level SP1, relative humidity
85 %
Preset warning level SP1, temperature
80 °C
Preset warning level SP1, viscosity ± 10 % (temperature compensated)
Preset warning level SP1, dielectric constant ± 15 % (temperature compensated)
Ambient Conditions
Nominal temperature range +20 .. +80 °C
Storage temperature -40 .. +90 °C
Fluid compatibility HLP mineral oils (HLP-D on request)
Esters: HEES, HETG
Seals: FPM.
-mark
EN 61000-6-1, EN 61000-6-2,
EN 61000-6-3, and EN 61000-6-4
Protection class to DIN 40050 IP 67
Other Data
Supply voltage (U
B
) 10 VDC to 36 VDC
Supply voltage ripple
5 %
Mechanical fitting G ¾ screw thread to DIN 3852 E
Torque value 30 Nm
Electrical connection
5-pin, M12x1 male
Reverse polarity and short circuit protection Provided
Housing material Stainless steel
Weight Approx. 205 g
“FS” (full scale) refers to the full dynamic range; “IV” refers to “initial value.”
* The best accuracy achievable when measuring relative humidity is strongly dependent upon the type of fluid or fluid
additive(s) involved. More precise information on this is available on request.
** The measurement accuracy of relative changes in viscosity and in dielectric constant vary according to the
application, the types of fluids involved, and the sensor’s own calibration. More detailed information on this is
available on request.
Przeglądanie stron 7